2L03 Commissioning Swing Shift
2L03 Commissioning Swing Shift
FSD Masking Tool: Successful System Masking for: BeamLossMonitor
MOLLER Install Update 03/10/2026
Hall A techs aided Rigging contractor with moving Pion donut assembly, from beam left side of hall to beam center, in front of the dump. Hall techs also aided Rigging contractors with moving Drift pipe from Test lab to Hall A.
OM06 Hall B ODH alarm
ARC7A has been enabled in alarm handler
"FC#2: Unlocked from "in" position"
The OPS DRVHs for the following cavities have been updated: 2L13-5
The OPS DRVHs for the following cavities have been updated:
2L13-5
Username: alexs
Date/Time: 03-10-26_18:38:18
RF Status: Good
Tuner Status: Good
Comment: DRVH=4.500 Cannot regulate above 4.5 MV/m, suspicious CRFP readback. log 4475490
Previous OPS DRVH: 7.2
Special Instructions:
2L13-5 unable to regulate above 4.5 MV/m
Cavity came back SOS'ing after returning SL to Power Permit. Initially suspected it might need tuning, but the detune angle looked fine, and soon after realized GASK was jumping back and forth between .5 and 10 V, while CRFP wasn't getting above 600 W and dropping to zero even with GASK railed. By walking down the gradient, it looks like the cavity can stay regulated at 4.5 MV/m, so leaving the drive high there for now. KCCU is stable at 1.24 A, so potentially a dying 2.5 W amp or RFCM going south.