Recent Log Entries
Follow-up Re: Hall A Target Change 10 uA of CW on Carbon Hole
Hall A Target Change 10 uA of CW on Carbon Hole
Hall A Target change request -
Target: Carbon Hole with 10 uA of CW
Figure 1: Target Ladder (Before Adjustments)
Figure 2: Ion Chambers adjusted
Figure 3: Target Ladder (After Adjustments)
Figure 4: Max Juice Settings
Figure 5: Hall A Operational Restrictions
Figure 6: Beam Authorization
Hall A No Target IC Calibration up to 10uA
Figure 1: Setup, raster on at nominal settings along with SBS dipole and correctors. BigBite also on at 750A.
XSOFs for no target, i.e. PolHe3 cell, dialed in.
Figure 2: Orbit. Figures 3-6: iCalibrate setup.
Figure 7: Plots during calibration.
Figures 8-14: Plots. Only 1H05, target upstream, has a better linear than log fit.
A-2023-09-25_0123_No_Target_Pass-4.hcd in ~mccops/IonChambers/HallA_2023/
iCalibrate: Hall A
Checking the Phase for 1L26-3
Using the 1A SLM to check the phase of 1L26-3. Sending about 5 uA of tune beam with the A laser to manually adjust the PSET value, and noting the beam displacement (checking for optimal/peak energy).
The cavity didnât require any adjustments in phase value.
Initial PSET value: -93.90
Final PSET value: -93.90
Initial PMES value: 39.5013
Final PMES value: 39.5068
Both PSET and PMES values measured in degrees.
Autolog save
HYST AREA 6-3-2 Cycles Magnets
HYST AREA Version 6-3-2 has successfully cycled the following:
MBD1C00V
MBD1C00AV
MBC1C01H
MBC1C02V
MBC1C03H
MBC1C04H
MBC1C05V
MBC1C06H
MBC1C07H
MBC1C07V
MBC1C09V
MBC1C11V
MBC1C13V
MBC1C15V
MBC1C18H
MBC1C18V
MBC1C20H
MBC1C20V
MBT1P01H
MBT1P04H
MBC1H01H
MBC1H01V
MBD1H04V
MBD1H04H
Total Time: 2.433 mins
HYST AREA 6-3-2 Cycles Magnets
HYST AREA Version 6-3-2 has successfully cycled the following:
MBD1C00V
MBD1C00AV
MBC1C01H
MBC1C02V
MBC1C03H
MBC1C04H
MBC1C05V
MBC1C06H
MBC1C07H
MBC1C07V
MBC1C09V
MBC1C11V
MBC1C13V
MBC1C15V
MBC1C18H
MBC1C18V
MBC1C20H
MBC1C20V
MBT1P01H
MBT1P04H
MBC1H01H
MBC1H01V
MBD1H04V
MBD1H04H
Total Time: 2.383 mins
Addressing Hall A 1H05 and 1H04B Ion Chamber Trips -- 8E02.X position adjustment silver bullet
When trying to deliver CW on Carbon Hole in Hall A, the first CW after a 2+ day down, found we couldn't deliver any appreciable beam current without 1H05, target upstream, and 1H04B, collimator, ion chamber trips. Switching to tune, could even trip these in tune-mode.
Adjusted some end of the line steering -- 1C20->1H04E -- around some to see if maybe we had a bad incoming angle that was causing us to scrape the collimator, tripping ICs. Was able to get some reduction in IC signals in tune-mode, but not enough for CW delivery without IC trips.
Injector Transmission with Only 30uA to xC Hovering Around ~70%
Operator SIGN IN
Operator SIGN IN
Experimenter Timesheet Summary: 24 Sep 2023 SWING
Experimenter Timesheet Summary: 24 Sep 2023 SWING
iCalibrate: Hall A
iCalibrate: Hall A
QuickPic - Hall A SBS and correctors turned on
QuickPic - Hall A Raster turned on at Values requested by Hall A
CREW CHIEF SIGN IN
THOMAS JEFFERSON NATIONAL ACCELERATOR FACILITY
ACCELERATOR OPERATIONS CREW CHIEF SIGN IN FORM
Date : 09-24-23
Time : 23:12:03
Day : Sunday
Shift : Owl
Crew Chief : Moser
Program Deputy: Benesch
Team : A
Arm : Howard
Designated SSO: Blackshaw
LERF Operator:
Operators : Blackshaw, Howard
This stamp certifies that the above Crew Chief has read/reviewed/received the following: